| Stock Code: 688115
| SC: 688115

SG2350 LCR Platform

SG2350 LCR Platform

  • Supports three testing excitation modes: open-circuit voltage, constant voltage,and constant current.
  • The maximum excitation level supported is 2 Vrms or 20 mArms, with DC bias capability.
  • Wide measurement range from 20 Hz to 2 MHz.
  • Basic impedance measurement accuracy better than 0.1%.
  • Supports up to 19 parameter measurements.
  • Fastest measurement speed of up to 5 ms.
  • Compact, modular design suitable for multi-channel parallel measurements.

Description

The SG2350 LCR Impedance Testing Platform is a precision LCR meter with a basic measurement accuracy of up to 0.1%. It supports various test excitation modes and offers a wide adjustable testing frequency range from 20 Hz to 2 MHz. It also provides continuous adjustable testing levels ranging from 0 to 2 Vrms or 0 to 20 mArms, along with a maximum adjustable DC bias of 2 V. This platform can be used to test various impedance parameters, ensuring precise measurements while achieving a rapid measurement speed of as fast as 5 ms. Its compact and modular design makes it a cost-effective choice for impedance parameter testing in production lines, components, materials, semiconductors, MEMS, and more.

sg2350 lcr platform

Product features

  • Supports three testing excitation modes: open-circuit voltage, constant voltage,and constant current.
  • The maximum excitation level supported is 2 Vrms or 20 mArms, with DC bias capability.
  • Wide measurement range from 20 Hz to 2 MHz.
  • Basic impedance measurement accuracy better than 0.1%.
  • Supports up to 19 parameter measurements.
  • Fastest measurement speed of up to 5 ms.
  • Compact, modular design suitable for multi-channel parallel measurements.

Product features

Primarily used for impedance parameter testing and measurement in components, materials, semiconductors, MEMS, and similar applications

sg2350 lcr platform

 

Specifications

The following indicators are based on open-circuit voltage mode

Measurement Frequency Range 20 Hz ~ 2 MHz
Step 10 Hz
Accuracy ±(100 ppm + 0.01 Hz)
Measurement Level Range 0 mVrms ~ 2 Vrms
Step 5 mVrms
Accuracy ±(10% + 1 mVrms)
DC Bias Voltage Range 0 ~ 2 V
Step 5 mV
Accuracy ±(0.1% + 2 mV)
Output Impedance 100 Ω (Typical Value)

 

Impedance Measurement Accuracy

Range Impedance Measurement Basic Accuracy (Medium speed, 1 KHz test frequency) Phase Measurement Basic Accuracy (Medium speed, 1 KHz test frequency)
1 Ω 2% 0.010 radians/ 0.573 degrees
10 Ω 0.3% 0.004 radians/ 0.230 degrees
100 Ω 0.1% 0.002 radians/ 0.115 degrees
800 Ω 0.1% 0.002 radians/ 0.115 degrees
3 KΩ 0.1% 0.002 radians/ 0.115 degrees
10 KΩ 0.15% 0.002 radians/ 0.115 degrees
30 KΩ 0.15% 0.002 radians/ 0.115 degrees
100 KΩ 0.15% 0.003 radians/ 0.172 degrees
300 KΩ 0.3% 0.003 radians/ 0.172 degrees
1 MΩ 1% 0.005 radians/ 0.286 degrees
3 MΩ 1% 0.010 radians/ 0.573 degrees
10 MΩ 4% 0.030 radians/ 1.719 degrees