Description
The SG2165 Precision Source-Measure Unit (SMU) platform is capable of four-quadrant operation,accurately providing voltage or current outputs while simultaneously measuring voltage, current, and resistance, among other functions. It integrates the capabilities of a six-and-a-half digit digital multimeter (DMM), a five-and-a-half digit precision voltage source, current source, electronic load, and pulse generator.It is feature-rich, compact in size, and comes with standard testing interfaces, making it highly suitable for integration into test fixtures. The SG2165 Source-Measure Unit platform is primarily used for IVR testing in industries such as semiconductors, sensors, and modules. It is customized for production line testing and provides efficient and cost-effective testing and measurement solutions for automated ICT and FCT in production lines.
Product features
- Combines the functions of 5 instruments (digital multimeter, voltage source, current source, electronic load, and pulse generator) into one unit.
- Typical output sources with measurement accuracy of 0.02%, supporting DC/pulse output modes.
- Pulse output mode with a minimum pulse width of 100 us and a rise time of 10 us.
- High-precision source with pA-level resolution and support for multi-channel parallel current expansion.
- Supports 2-wire and 4-wire resistance measurements.
- Supports SCPI driver and commands.
I-V Testing Performance
Application Scope
Primarily used for IVR testing in industries such as semiconductors, sensors, and modules.
It is customized for production line testing and provides efficient and cost-effective testing and measurement solutions for automated ICT and FCT in production lines.
Specifications
Model | SG2165 | ||||
Specifications | 20 W | ||||
Channels | 1 | ||||
Quadrants | 4 | ||||
Voltage Accuracy | Range | Setting Resolution | Setting Accuracy | Measurement Resolution | Measurement Accuracy |
200 mV | 2 uV | 0.02%+375 uV | 15 nV | 0.02%+375 uV | |
2 V | 20uV | 0.02%+600 uV | 150 nV | 0.02%+600 uV | |
20 V | 200uV | 0.02%+5 mV | 1.5 uV | 0.02%+5 mV | |
Current Accuracy | Range | Setting Resolution | Setting Accuracy | Measurement Resolution | Measurement Accuracy |
100 nA | 1 pA | 0.06%+50 pA | 10 fA | 0.06%+50 pA | |
1 uA | 10 pA | 0.03%+700 pA | 100 fA | 0.03%+700 pA | |
10 uA | 10 0pA | 0.03%+5 nA | 1 pA | 0.03%+5 nA | |
100 uA | 1 nA | 0.03%+60 nA | 10 pA | 0.03%+60 nA | |
1 mA | 10 nA | 0.03%+300 nA | 100 pA | 0.03%+300 nA | |
10 mA | 100 nA | 0.03%+6 uA | 1 nA | 0.03%+6 uA | |
100 mA | 1 uA | 0.03%+30 uA | 10 nA | 0.03%+30 uA | |
1 A | 10 uA | 0.05%+1.8 mA | 100 nA | 0.05%+1.8 mA | |
3 A (≤6 V ) | 100 uA | 0.4%+8 mA | 1 uA | 0.4%+8 mA | |
Output | Voltage Source | Range | Typical Output Settling Time | Test Conditions | |
200 mV | <20 us | Voltage Change Time at No Load, 10%-90% | |||
2 V | <50 us | ||||
20 V | <200 us | ||||
Current Source | 100 nA | <10 ms | Current Change Time at Output Short Circuit,10%-90% | ||
1 uA | <5 ms | ||||
10 uA | <5 ms | ||||
100 uA | <1 ms | ||||
1 mA | <300 us | ||||
10 mA | <20 us | ||||
100 mA | <20 us | ||||
1 A | <20 us | ||||
Voltage Ripple and Noise (10 Hz-20 MHz) | 3 mVrms, 20 V range |