Description
The SG2163 Vector Network Analyzer (VNA) is a four-port RF measurement instrument capable of providing measurements of RF signal transmission characteristics and reflection characteristics. This product consists of a main unit and a control and display interface based on the Windows operating system, with data transmission utilizing a Gigabit Ethernet interface.It finds wide applications in RF research,testing, and production manufacturing in fundamental industries and fields such as microwave devices,materials science, and electronic communications.
Product features
- Host unit with four RF test ports
- Testing dynamic range of 120 dB or more
- Low background noise down to -125 dBm/Hz
- User interface based on the Windows system
- Measurement time per point better than 70 microseconds
- Supports RF signal testing up to 8.5 GHz
- Supports measurements such as S-parameters, Smith charts, and frequency response
Application Scope
Specifications
Number of test ports | 4 (2 or more ports are custmizable) |
System impedance | 50 ohm |
Connector of test ports | SMA female (SMA or N is custmizable) |
Damage level | +26 dBm , 35 VDC |
Frequency range | 100 KHz to 8.5 GHz (custmizable) |
Frequency resolution | < 1 Hz |
Power range | -55 dBm to +5 dBm |
Power accuracy | ±1.5 dB |
Power resolution | 0.05 dB |
Dynamic range | 120 dB |
Measurement bandwidth(IFBW) | 10 Hz to 1 MHz |
Harmonic distortion | -20 dBc |
Non-harmonic spurious | -20 dBc |
Level for 0.1 dB compression | >+5 dBm |
Noise floor | -125 dBm/Hz |
Trace noise (IFBW 1 KHz) | 0.003 dBrms / 0.05 degrms |
Measurement Accuracy of transmission | ±0.2 dB @ 0 dB to +10 dB ±0.1 dB @ -10 dB to 0 dB ±0.2 dB @ -40 dB to -10 dB ±1.0 dB @ -65 dB to -40 dB ±2.0 dB @ -85 dB to -65 dB |
Measurement Accuracy of reflection | ±0.4 dB @ -15 dB to 0 dB ±1.0 dB @ -25 dB to -15 dB ±3.0 dB @ -35 dB to -25 dB |
Directivity | 45 dB |
Source match | 40 dB |
Load match | 45 dB |
Transmission tracking | ±0.1 dB / ±1° |
Reflection tracking | ±0.1 dB / ±1° |
Number of measurement points | 2 to 16001 |
Measurement time per data point | 70 us (typical value) |
Power | 12 VDC ± 5% |
Current | 3.4 A, maximum |
Power consumption | 36 VA (typical value) |
Dimensions | 230 mm * 370 mm * 70 mm (with chassis or modularized is custmizable) |
Weight | 5.0 kg |
Operating temperature range | 0 °C to + 50 °C |
Maximum altitude | 2,000 m (at 25 °C ambient temperature) |
Relative humidity range | 10% to 90% R.H. , no condensation |